Analytical Instrumentation Lab: FINAL

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/104

flashcard set

Earn XP

Description and Tags

All Questions from the slides for the last 3 slides

105 Terms

1
New cards

What is electrophoresis?

a technique used to separate macromolecules such as DNA or protein

2
New cards

What type of gel serve as a molecule sieve in electrophoresis?

Agarose or Polyacrylamide Gels

3
New cards

How does gel electrophoresis use charge to separate molecules?

positive and negative charges. Positive charged molecules will migrate toward the negative electrode and vice versa

4
New cards

What is a buffer in gel electrophoresis?

A solution that conducts electricity that is poured into the electrophoresis chamber to allow the electric current to flow from the cathode, through the buffer, to the anode.

5
New cards

What are the functions of protiens? (6)

Enzyme catalysis, Metabolic Regulation, Binding and transport of small molecules (i.e.,Transferrin, Ceruloplasmin and Hemoglobin), transcription factors, antibodies, structural proteins in cell walls

6
New cards

What 2 amino acids are responsible for proteins negative charge?

Glutamic acid and aspartic acid

7
New cards

What 2 amino acids are responsible for a proteins’ positive charge?

Lysine and arginine

8
New cards

What is the isoelectric point in a protein? Why does this matter in gel electrophoresis?

The pH at which the protein has no net charge which means a protein will not migrate in an electric field at its’ isoelectric point, so you can alter the direction and extent of a protein’s migration by using acidic, neutral, or basic buffers during electrophoresis

9
New cards

What determines a protein’s charge? (2)

amino acid composition and pH conditions

10
New cards

Proteins having _____ charge than others of the same shape and size will travel faster

more

11
New cards

Proteins that are tightly folded will move through the gel _____ than a protein that is more loosely folded

faster

12
New cards

Define native protein

protein in their normal biologically active forms

13
New cards

What can cause a protein to become denatured? (3)

detergents, extremes in pH, and organic solvents

14
New cards

What defines the molecular weight of a protein? (2)

the number and type of amino acids in the polypeptide chain

15
New cards

What are some factors that affect the electrophoretic migration rates? (4)

the amount and sign of the charge, the side and shape of the native protein

16
New cards

What type of proteins travel the furthest? Which travel the shortest distance? (Gel Electrophoresis)

Small high positive charged proteins travel the furthest, large low positive molecules travel the least

17
New cards

What protein in the human body has one of the fastest migration rates?

albumin

18
New cards

What protein in the human body has one of the slowest migration rates?

gamma globulins (antibodies)

19
New cards

What is SDS-PAGE protein electrophoresis?

A type of electrophoresis that separates proteins by molecular weight biases SDS in the buffer that unfolds the protein and coats it in a negative charge

20
New cards

What are the 2 chemicals in SDS-PAGE Electrophoresis that help it separate protiens by molecular weight only?

SDS and BME

21
New cards

What does SDS do in SDS-PAGE Electrophoresis? What does BME do?

A chemical contained in the sample buffer and the gel that unfolds the protein and coats it with a negative charge; DME is a reducing agent that breaks the disulfide bonds in the protein

22
New cards

SDS-PAGE Electrophoresis makes a protein’s charge (proportional/ inversely proportional) to it’s _________?

proportional to it’s molecular weight

23
New cards

What does the PAGE in SDS-PAGE Electrophoresis stand for? What is it?

Poly Acrylamide Gel Electrophoresis; the type of gel used that can control pore size and sets chemically via polymerization rather than thermally and it run vertically rather than horizontally with a stacking gel on top of a resolving gel

24
New cards

What is the stacking gel versus the resolving/ running gel in SDS-PAGE electrophoresis?

The stacking gel is at pH 6.8 on top of the resolving gel which has large pores and the wells. The resolving gel is at pH 8.8 and has small pores that resolves the proteins according to size

25
New cards

What does it mean to have a discontinuous buffer? (as is the case for SDS-PAGE electrophoresis)

the buffer in the tank is a different pH than the buffer used to make the gel which creates a gradient where all proteins are focused into one single band in the large pores of the snacking gel and resolve in the smaller pores of the resolving gel which provides high resolution and good band definition

26
New cards

What are some factors that will affect the separation of protein bands?

speed protein migrates, proteins sensitivity to pH, proper voltage (since too high voltage will cause the proteins to migrate too fast or melt the gel), and increasing the ionic concentration of your buffer that will decrease the speed of migration

27
New cards

What is the Bio-Rad Mini-PROTEAN Tetra Cell System?

A mini cell system for 1-D vertical gel electrophoresis that runs SDS-PAGE on 1-4 mini gels in 30 minutes

28
New cards

What defines wether the chromatography is gas or liquid?

Wether the mobile phase is either a gas or a liquid

29
New cards

What is Gas-Liquid Chromatoraphy?

The mobile phase is gas and the stationary phase is liquid

30
New cards

What phase of chromatography is the sample introduced into first?

The mobile phase

31
New cards

What is the mobile phase usually in Gas Chromatography?

An inert gas like Helium that is released from a compressed gas cylinder

32
New cards

What type of mixtures are used for gas chromatography?

organi liquids

33
New cards

What form must the mixtures be the injected into the gas chromatographer?

in the gas phase/ rapidly evaporated immediately upon injection

34
New cards

Why must the components of a Gas Chromatography device be kept at an elevated temperature?

To ensure the sample is instantly converted into gas and to keep the gas belittle and moving to avoid condensation

35
New cards

Where is the stationary phase physically located within the gas chromatography device?

bound to the interior wall of the capillary tube

36
New cards

What factors determine the separation of components in a gas chromatography device?

The solubility and vapor pressure/ how well the mixture components dissolve in the liquid stationary phase and how well the mixture components evaporate/ stay in the mobile phase

37
New cards

Define vapor pressure

A substance’s tendency to be in the gas phase at a given temperature

38
New cards

Substances with a higher vapor pressure will stay in the gas phase _____ and move through the column _____ than substances with lower vapor pressure

longer, faster

39
New cards

Substances with higher solubility will do what in the stationary phase? How fast do they move through the column

They will stay in the stationary phase/ dissolve on the column and move through more slowly than substances with low solubility

40
New cards

What are the parts of the injection port of a gas chromatography device?

a small glass lined cylinder with a rubber septum to accommodate a syringe injection of a small volume of liquid

41
New cards

What can you do to help increase the resolution of chromatograms? Why?

change the column length due to more interaction of the sample with the stationary phase

42
New cards

What is the more common style of capillary tube in GC? Why?

Open/ capillary tubular capillaries since they have better resolving power

43
New cards

What is the difference between open/ capillary tubular columns and packed columns in GC?

open tubular columns allow little resistance to gas flow where packed columns need higher gas pressure to push the samples through and open columns have better resolving power

44
New cards

What does it mean for a GC separation to be “partition type”?

It means they utilize a liquid stationary phase

45
New cards

How is the stationary phase bonded to the capillary tube in GC? What do silanes do?

Capillary tubes are made of fused silica (a synthetic quartz material) and the inside is coated with a polyamide that undergoes “silylation” that exposes silanol groups that bind to the stationary phase to form silanes. These silanes complex and result in polysilane liquids that are fixed via chemical bond to the artificial quartz. Silanes give the column a polar or non-polar property.

46
New cards

What is the liquid stationary phase bonded to in a packed column GC?

diatomaceous earth inside the column

47
New cards

What is temperature programming in GC?

For more complicates separations a single linear increase from a low to high temperature is needed as the substance moves through the column

48
New cards

What happens when the temp is too high in GC? What about too low?

Too high is poor resolution of high vapor pressure components, too low makes the retention times of low vapor pressure components too long

49
New cards

How does the GC detect substances/ when does it create peaks?

It creates peaks when gas other than the carrier gas exits the column proportional to the amount of analyte present

50
New cards

What is the most common and most widely applicable type of GC detector?

Flame Ionization Detector (FID)

51
New cards

How does an FID work?

As analyte gasses escape, they are burnt by an H flame and small fractions of the molecules are fragmented and these bonds are drawn to a - or + electrode or burner head and the Curren change between the electrical circuit shows up as a peak on a chromatogram

52
New cards

What is the universal detector in GC?

Thermal conductivity detector

53
New cards

How does a thermal conductivity detector work in GC?

detects any components other than the carrier gas by comparing the change in reference flow of the carrier gas using a Wheatsone Bridge circuit to measure resistances as a function proportional to temperature

54
New cards

What type of carrier gas is usually selected for GC? name 5 examples

inert gas like helium, nitrogen, argon, hydrogen, or air

55
New cards

What is the difference between split and split less injector valves? Why use a split valve?

A split injector vents some of the vaporized sample through a split vent so that only a small fraction of the sample actually goes through the column. Split injection prevents column overload and is good for low sensitivity high concentration samples.

56
New cards

How is there an “Internal Standard” (IS) in chromatograms?

All standards and samples are spiked with a constant amount of a substance, so it serves as a reference for peak size measurements.

57
New cards

How do you use the Internal Standard method to find the amount of a substance in GC?

measure peak sizes of both the internal standard peak and the analyte peak, then divide the analyte peak area by the IS peak area. The area ratio is then plotted in a linear graph of area ratio vs concentration of analyte

58
New cards

What does it mean when there are diminished peak sizes on a chromatogram? What could this be caused by? (GC) (6)

there is reduced sensitivity which usually means less sample was injected than assumed or the the detector is sending smaller electrical signals. Ir could also be due to a leaking syringe, a worn septum, a leak in the pre or post column connections, or a contaminated detector

59
New cards

What is peak “fronting” or peak “tailing” in GC and what could it be caused by?

When the chromatogram’s peaks lean in a certain direction instead of being straight up and down. This could be due to too large injection volume, too slow of syringe injection, or not fully penetrating the septum. It could also be due to decomposition of thermolabile components or contamination of the injection port or column.

60
New cards

What are 3 things that could cause unwanted changed in the carrier gas flow rate? (that affected retention times)

leaks in the injection port, low pressure from an empty or almost empty fas tank, faulty gas pressure regulator/ flow controller

61
New cards

What can cause altered retention times in GC?

Changes in carrier gas flow rate or changes in temperature

62
New cards

What can cause unwanted changes in temperature in GC? (Altered retention times) (4)

Faulty temp controller, improperly set temp program, too short a cool down prior to the next injection, column. overload or decomposition of stationary phase

63
New cards

What is Baseline Drift in a chromatogram? What causes it?

When the graph’s peaks don’t start at the bottom and continually rise up which ossuary when a new column has not been “baked out” overnight before use. This could also be due to detector temps not reaching equilibrium value/ being faulty.

64
New cards

What are baseline perturbations in Chromatograms?

deviations in the system/ baseline like irregular spikes or regular but unexpected spikes

65
New cards

What could cause irregular spikes on a chromatogram?

detector contamination or interference from nearby electrical equipment

66
New cards

What could cause regular unexpected spikes on a chromatogram?

Condensation in the column causing the carrier gas to pulse, electrical pulses, or pulses in carrier gas flow due to a a faulty flow valve/ pressure regulator

67
New cards

What are 4 things you can try when you get unexpected spikes on your chromatogram?

perform a “bake out”, replace the septa, carrier gas, and/or column

68
New cards

What are 6 factors that affect sample separation?

boiling point, polarity, column temperature, affinity of sample for column, carrier gas flow rate, and length of column

69
New cards

(HPLC) What is HPLC?

A separation technique with a liquid mobile phase and is a high performance option that can identify many different biological compounds when the sample and moving phase interact chemically

70
New cards

(HPLC) Compare HPLC with GC

HPLC has a liquid mobile phase while GC has a gas mobile phase, in HPLC, the samplee interacts with the mobile phase, it is run at room temperature, both have the same injector design

71
New cards

(HPLC) What is the mobile phase of HPLC usually (3)?

Acetonitriles, methanol, or water

72
New cards

(HPLC) Where is the mobile phase contained?

In a glass bottle with a vented cap and a filter inside of a tube that draws it up towards the pump via vacuum

73
New cards

(HPLC) What does the liquid mobile phase need to undergo? (2)

Needs to be sent through a filter in the tubing to remove particles and needs to be “degassed” to avoid bubbles using a degasser since it needs to be HPLC/ solvent grade (pure)

74
New cards

(HPLC) What are the 2 mobile phase elution methods that are used to elute mixture components from the stationary phase?

Isocratic and gradient

75
New cards

(HPLC) What is Isocratic mobile phase elution method? What is Gradient?

one mobile phase composition is used for the entire seperation while in gradient, the mobile phase composition is changed gradually in the middle of the run (similar to temp programming in GC)

76
New cards

(HPLC) What is the difference between rapidly eluting and slowly elution compounds?

rapidly eluting components are highly soluble in the mobile phase, but insoluble in the stationary phase while vise versa in slowly eluting components

77
New cards

(HPLC) How can renting times and resolution be altered (on purpose)?

By changing the composition of the mobile phase using gradient elution

78
New cards

(HPLC) What type of injector is used? Why?

A loop injector with a 2 position valve that directs the mobile phase along 1 of 2 paths where 1 path is the sample loop and the other path bypasses the loop and continues into the column. This is due to how high the pressure of the mobile phase is since it could destroy a regular rubber septum

79
New cards

(HPLC) How are solutions injected into the sample loop?

They are filled to overflowing every time and the sample loop is manufactured to deliver a particular volume, there are also automated injectors in devices that handle a high number of samples with a robotic needle aspirating samples from a glass vial

80
New cards

(HPLC) What are 4 types of columns that are unique to HPLC?

adsorption, partition, ion exchanges and size exclusion

81
New cards

(HPLC) What is the most common column used for HPLC? What are the subtypes (3)?

Partition is the most common and can be further classified as normal phase, reserve phase, or bonded phase

82
New cards

(HPLC) What is a normal phase partition HPLC column? What are the mobile phases usually?

consists of a non-polar mobile phase in combination with a polar liquid stationary phase; mobile phase is usually Hexane, Cyclohexane, Carbon tetrachloride, chloroform, benzene, and toluene

83
New cards

(HPLC) What is a reserve phase portion HPLC column? What are they usually named? What type of mobile phases are usually used?

Utilized a polar mobile phase with a non-polar stationary phase. Column names indicate the length of carbon chains; Mobile phase is usually water, methanol, acetonitrile, and acetic acid buffered solutions

84
New cards

(HPLC) What are absorption columns?

A classification of HPLC where the highly polar silica particles are expose (not absorbed or bonded to a liquid phase) and are made of aluminum oxide particles. This uses normal-phase column mobile phases

85
New cards

(HPLC) What are Ion Exchange columns? What is the mobile phase?

Columns consist of solid resin particles that have positive and/or negative ionic bonding sites where ion exchange occurs with the mobile phase. The mobile phase is always a pH buffered solution

86
New cards

(HPLC) What are size-exclusion columns?

One where picture components are separated by size via molecules interacting with various pore sizes on the surface of the porous polymeric particles

87
New cards

(HPLC) What is the typical size of the stationary phase particles and why does this matter?

The typical size is 5um and the smaller the pore size the better the resolution because they provide a greater surface area for separation

88
New cards

(HPLC) What is “ultra-performance liquid chromatography”?

Chromatography with a stationary phase particle size smaller than 2um that provides greatest surface area for seperation'/ highest resolution

89
New cards

(HPLC) What factors into your column selection?

the mixture components should have a high affinity for the stationary phase, separation parameters should be adjusted, using partition chromatography for highly polar mixtures and adsorption chromatography for very non-polar mixtures

90
New cards

(HPLC) What type of detectors can be used for HPLC (3)?

spectrophotometers, fluorometers, and refractometers, these methods are more conventional

91
New cards

(HPLC) What is a UV Absorption detector? How does it work? What is a limitation?

A UV spec that measures flowing rather than a static solution. as the sample passes through the flow cells, the absorbance changes and a chromatogram is generated. Can only be used on samples that absorb UV light and the mobile phase should not absorb UV light

92
New cards

(HPLC) What is a Diode Array Detector?

A UV spectrophomoters with a flow cell. Light from the UV source passes through a flow cell and is dispersed via a grating. The entire UV absorption spectrum can be measured as each individual components elutes. It allows for a definite qualitative analysis to be made based on the spectrophotometric fingerprint of the sample

93
New cards

(HPLC) How does a fluorescence detector work? What is a benefit and a downside?

individually measures the fluorescence intestines of the mixture components as they elude from the column; a benefit is that it is selective and sensitive, but it is not universal

94
New cards

(HPLC) What is a refractive index detector? What are some benefits and drawbacks?

A detector that consists of a refractometer and a flow cell and focuses a light beam on a diode array. Takes the refractive index of the pure mobile phase as a reference point and when mixture components elute, the refractive index changes giving an electrical signal; benefit is that it is a universal detector, but it is not sensitive since output is subject to temperature effects and mobile phase composition during gradient elution

95
New cards

(HPLC) What does qualitative analysis depend on? What about quantitative?

Qualitative depends solely on retention time/ Quantitive depends on the peak size based on a series of standards

96
New cards

(HPLC) What are the most common causes of issues in HPLC?

high or low pressure, system leaks, worn injector parts, air bubbles and/or blocked inline filters.

97
New cards

(HPLC) What can cause unusually high pressure?

clogged in-line filter, injector blockage, mismatched mobile and stationary phases, flow rate too high

98
New cards

(HPLC) What can cause unusually low pressure?

Suggests a leak in the system, check all joints for leaks. Can occur within the pump seal, insert, various fittings and joints around the column and detector.

99
New cards

(HPLC) What do air bubbles do in HPLC? (3) Why?

Air pockets in the pump can cause slow or no pressure or flow, erratic pressure and changed in the Tr, this is because they result in decreased contract with the stationary phase leading to shorter Tr and decreased resolution

100
New cards

(HPLC) How can you get rid of air bubbles in the HPLC?

increasing flow rate or restricting then releasing the post detector flow in order to increase the pressure to blow out the bubbles