Electrophoresis
An analytical technique used to separate macromolecules such as DNA or proteins using an electric field and a gel matrix acting as a molecular sieve
Gel (Electrophoresis)
A porous matrix held in a casting tray that acts as a molecular sieve to slow down particle migration toward oppositely charged electrodes
Comb
A multi-toothed tool placed in slots of the casting tray before the gel solidifies to form sample loading wells
Wells
Small cavities created in the gel by the teeth of the comb that provide a place to load sample solutions
Buffer (Electrophoresis)
A conductive solution covering the gel that allows electric current to flow from the cathode to the anode
Native Protein Migration Factors
The net charge, shape (conformation), and size (molecular weight) of a native protein that dictate its electrophoretic migration speed
SDS (Sodium Dodecyl Sulfate)
A detergent in sample buffer and gel that unfolds proteins and coats them with a uniform negative charge proportional to their molecular weight
BME (Beta-Mercaptoethanol)
A reducing agent used in SDS-PAGE sample buffer to break intra- and inter-chain disulfide bonds
Stacking Gel
The upper layer of polyacrylamide gel at pH 6.8 with large pores that focuses loaded proteins into a single thin starting band
Resolving Gel
The lower layer of polyacrylamide gel at pH 8.8 with small pores that resolves proteins strictly according to their molecular mass
Discontinuous Buffer System
A buffer system where the tank buffer pH differs from the gel buffer pH, creating a gradient that focuses proteins before resolution
Albumin Migration Rate
Fastest migrating human serum protein due to its relatively small molecular weight (~66.5 kDa)
Gamma Globulins Migration Rate
Slowest migrating serum proteins due to their large molecular weight (146,000-970,000 kDa)
High-Performance Liquid Chromatography (HPLC)
An analytical separation technique using a high-pressure pump to drive a liquid mobile phase through a stationary phase column
HPLC Mobile Phase
The liquid solvent (e.g. , acetonitrile, methanol, water) that carries sample components through the HPLC column
Vented Cap (HPLC Reservoir)
A cap on the solvent reservoir that allows pressure equalization with the atmosphere while preventing contamination
Sinker Filter
A weighted metal filter at the bottom of the draw tube that keeps tubing submerged and removes particulates from the solvent
HPLC Degasser
A device that removes dissolved gases from mobile phase solvents to prevent air bubble formation that disrupts pump flow and detection
Isocratic Elution
An HPLC elution method where the mobile phase composition remains constant throughout the entire chromatographic run
Gradient Elution
An HPLC elution method where the mobile phase composition is gradually altered during the run to optimize separation speed and resolution
Sample Loop Injector
A multi-port valve used in HPLC to inject precise, reproducible sample volumes under high pressure without using a rubber septum
Normal-Phase HPLC
A liquid chromatography mode using a polar stationary phase and a non-polar mobile phase, eluting non-polar compounds first
Reverse-Phase HPLC
A liquid chromatography mode using a non-polar stationary phase and a polar mobile phase, eluting polar compounds first
Diode Array Detector (DAD)
An optical HPLC detector that measures the entire UV-Vis absorption spectrum simultaneously for each eluting peak
Gas Chromatography (GC)
An instrumental separation technique that volatilizes liquid samples and sweeps them through a heated column using an inert carrier gas
GC Mobile Phase (Carrier Gas)
An inert gas (typically helium, nitrogen, argon, hydrogen, or air) that propels vaporized sample components through the GC column
Vapor Pressure (GC)
A substance's tendency to exist in the gas phase at a given temperature; compounds with higher vapor pressure move faster through the column
Solubility (GC)
A substance's tendency to dissolve in the liquid stationary phase; compounds with higher solubility move slower through the column
Split Injection
A GC injection mode where the split valve is open to vent most of the vaporized sample to waste, preventing capillary column overloading
Splitless Injection
A GC injection mode where the split valve is closed to direct virtually all vaporized sample onto the column, used for trace analysis
Capillary Column
A long (up to 300 ft) open tubular column coiled inside a GC oven with stationary phase bound to its inner wall, providing high resolution
Flame Ionization Detector (FID)
A common, destructive GC detector sensitive to hydrocarbons that burns sample components in a hydrogen flame to generate an electrical signal
Thermal Conductivity Detector (TCD)
A universal, non-destructive GC detector that measures changes in thermal conductivity using a Wheatstone bridge circuit
Internal Standard (IS)
A constant reference substance spiked into all standards and samples to normalize peak area ratios and compensate for injection technique variations
Fatty Acid Methyl Esters (FAMEs)
Volatile ester derivatives produced by reacting fatty acids with methanol to enable analysis by gas chromatography