Comprehensive Study Notes on Flow Cytometry and Laboratory Applications

Introduction to the BD FACS Caliber

  • Instrument Identity: The specific machine being discussed is the BD FACS Caliber.     * BD: Stands for Becton and Dickinson, the company that manufactures this particular line of instruments.     * FACS: Stands for Fluorescence Activated Cell Sorter.

  • Terminology: While "flow cytometer" is the technical term for the technology, over time, the term FACS machine has become interchangeable with "flow cytometer."

  • Product Line: BD produces a variety of these machines including the FACS Canto, FACS Aria, and others.

Sample Requirements for Flow Cytometry

  • Requirement of Suspension: Any particle or cell can be studied as long as it can be put into suspension (floating in a liquid).

  • Size Constraints:     * The machine's operational range is approximately 0.2μm0.2\mu\text{m} to 50μm50\mu\text{m}.     * While point 0.2μm0.2\mu\text{m} is the technical limit, a comfortable range for this machine is 0.5μm0.5\mu\text{m} to 50μm50\mu\text{m}.

  • Processing Non-Suspended Samples:     * Tumors: Must be broken up into individual cells.     * Plant Leaves: Must be processed into individual cells or individual nuclei before analysis.

  • Study Subjects: Examples of items that can be analyzed include:     * Allergies     * Chromosomes     * Red blood cells     * Protozoans

Information Derived from the Flow Cytometer

  • Without Staining: The machine provides two primary pieces of information based on light scatter:     * Relative Size: Determined by Forward Scatter (FSCFSC).     * Internal Complexity: Determined by Side Scatter (SSCSSC). Complexity refers to extensions, granules, or grooves on the cell.

  • With Staining: If stained with a fluorescent dye or if the sample glows naturally (autofluorescence), the machine measures Fluorescence Intensity.

The Fluidics System

  • Purpose: To transport cells or particles in a fluid stream to the laser beam and position the sample precisely in the center of that beam.

  • Operational Process:     * Air passes through an air filter to remove dust and particles.     * A pump moves the air across a regulator set to exactly 4.5 pounds per square inch4.5\text{ pounds per square inch} (psipsi).     * The pressurized air enters the Sheath Tank, which contains Sheath Fluid (a saline solution).     * The pressure shoves the liquid out of the tank, through a sheath filter (to remove bubbles or debris), and into the flow cell.     * The fluid and sample eventually exit into a waste tank.

  • Hydrodynamic Focusing: Due to the shape of the flow cell and the speed of the fluid, the sample is kept in a very tight stream as it passes the laser.

  • Pressure Settings:     * Low Pressure: The sample moves slowly. This is used for rare events or highly accurate counts (e.g., catching specific mitotic cells). The machine is sensitive enough to pick up a single glowing protozoan in a gallon of seawater.     * High Pressure: The stream is wider, and more cells pass at once. This is used for large volumes of patient samples where only percentages are needed, as some events may be missed.     * Medium Pressure: An intermediate setting exists as well.

The Optics System

  • Excitation Component:     * Uses lasers as the light source. The machine described has a 488 nm488\text{ nm} laser and a 635 nm635\text{ nm} laser.     * Lenses focus the laser beam onto the cell as it passes through the flow cell.

  • Collection Component:     * Consists of mirrors and filters that route emitted or scattered light to specific detectors.

  • Filter Types:     * Long Pass (LP): Allows wavelengths longer than a specific threshold to pass (e.g., LP 500LP\,500 allows ≥500 nm\ge 500\text{ nm} and reflects shorter waves).     * Short Pass (SP): Allows wavelengths shorter than a threshold to pass (e.g., SP 500SP\,500 allows ≤500 nm\le 500\text{ nm} and reflects longer waves).     * Band Pass (BP): Allows a specific range to pass. A BP 500/50BP\,500/50 allows a range of 50 nm50\text{ nm} centered at 500 nm500\text{ nm} (475 nm475\text{ nm} to 525 nm525\text{ nm}).

  • Detectors (Photomultiplier Tubes - PMTs):     * FL1FL1: Uses a 530/30 nm530/30\text{ nm} filter. Detects green light, commonly for dyes like FITC.     * FL2FL2: Uses a 585 nm585\text{ nm} filter. Detects yellow-orange light, commonly for dyes like PE.     * FL3FL3 and FL4FL4: Additional detectors for more colors.

Light Scattering and Fluorescence Mechanics

  • Forward Scatter (FSCFSC): Light scattered in a forward direction relates to the surface area or relative size of the particle.

  • Side Scatter (SSCSSC): Light scattered to the side (90∘90^{\circ} angle) relates to cell granularity and internal complexity.

  • Fluorescence Intensity: The brightness recorded is directly proportional to the number of binding sites. If a cell is twice as bright (higher fluorescence), it has twice as many binding sites/proteins labeled.

The Electronics System

  • Function: Converts the light signal (photons) into a proportional electronic signal (a number) for computer analysis.

  • Signal Processing: PMT detectors convert photons to electronic pulses; dim light creates a low signal/number, while bright light creates a high signal/number.

  • FCS File Format: Data is saved in the International FCS (Flow Cytometry Standard) format, also known as List Mode Data.     * It records measurements on a cell-by-cell basis for each parameter (FSCFSC, SSCSSC, FL1FL1-FL4FL4).     * For clinical integrity, the numbers in these files cannot be changed or edited in the standard software to prevent falsification of patient data.

Applications and Data Analysis

  • Blobology: The visual clustering of dots on a scatter plot (FSCFSC vs. SSCSSC) where different cell types (neutrophils, monocytes, lymphocytes) form distinct groups based on their specific size and complexity.

  • Cell Cycle Analysis: Using DNA dyes like Propidium Iodide (PI) to measure DNA content. S-phase cells (synthesizing DNA) show higher fluorescence than G1 cells. This is used in anti-cancer drug research to monitor tumor growth inhibition.

  • Clinical Diagnostics:     * HIV Monitoring: Counting CD4+CD4+ T-helper cells to check drug effectiveness.     * Blood Banks: Checking for residual white blood cells in blood products to prevent recipient immune responses.     * Prostate Cancer: Searching for new, more accurate biomarkers beyond PSA.

  • Environmental and Industry:     * Water Industry: Checking for Cryptosporidium (famous for the 1993 Milwaukee outbreak affecting ≈400,000\approx 400,000 people). Cryptosporidium is chlorine-resistant, requiring ozonation.     * Agriculture: Checking sperm viability for artificial insemination.     * Wine Industry: Checking the viability and health of yeast before grape fermentation.

Questions & Discussion

  • Question (Student): Can the machine collect the cells after analysis?     * Response: This specific machine currently sends everything to waste. It can be upgraded with a sorter (≈$30,000\approx \$30,000 attachment) that uses a valve to "click" and drop specific cells into tubes. Modern high-end sorters (≈$350,000\approx \$350,000) use electrical charges to pull cells into tubes, which preserves a higher concentration of cells than the gravity-drop method. Charge-based sorting is avoided for dangerous samples like anthrax to prevent aerosolization.

  • Question (Student): Can it detect blue dyes/labels?     * Response: This specific machine lacks a laser/detector for the blue spectrum (e.g., DAPI). While other machines can do this, for this unit, a blue-labeled cell would look the same as an unlabeled cell in the fluorescence detectors.

  • Question (Student): What is Phase Contrast microscopy?     * Response: Developed by Fritz Zernike (Nobel Prize winner), it uses differences in the refractive index of cellular structures to create levels of gray. This allows the visualization of live, unlabeled cells and their internal organelles without killing or staining them.

Laboratory Session: N2a Cell Differentiation

  • Context: The students are preparing the differentiation of N2a cells (Day 1).

  • Reagents: Differentiating media (MEM supplemented with 1% FBS1\%\,FBS) and Retinoic Acid (RA).

  • Retinoic Acid Concentration: Target final concentration is 10 μmol dm−310\,\mu\text{mol}\,dm^{-3}.

  • Protocol Steps discussed:     * Warm the media.     * Wash cells with PBS (typically 1 ml1\text{ ml} for a 12-well plate).     * The wash is performed twice for 3−5 minutes3-5\text{ minutes}.     * Discard PBS into waste, then add the differentiated media containing RA.     * Incubate to continue the differentiation process.